Electric Vacuum Muffle Furnace Ovens For Laboratory Annealing Decarburization

Brand Name: Chitherm
Model Number: HWF160-10NH
MOQ: 1
Price: Negotiable
Delivery Time: Customized
Payment Terms: Customized
Detail Information
Place of Origin:
China
Range Of Applications:
Industrial
Type:
Electric Holding Furnace
Usage:
Ceramic Sintering
Fuel:
Electric
Atmosphere:
Vacuum/Nitrogen/Hydrogen
Effective Chamber Dimensions:
500*500*650mm(W*H*D)
Transport Package:
Wooden Packaging
Specification:
1650*1950*2000mm(W*H*D)
Trademark:
Chitherm
Origin:
China
HS Code:
8514101000
Production Capacity:
50 Sets/Year
Packaging Details:
Customized
Supply Ability:
Customized
Highlight:

decarburization Vacuum Muffle Furnace

,

Laboratory Vacuum Muffle Furnace

,

decarburization muffle ovens laboratory

Product Description
Electric Vacuum Muffle Furnace Ovens For Laboratory Annealing Decarburization
Product Attributes
Attribute Value
Range of Applications Industrial
Type Electric Holding Furnace
Usage Ceramic Sintering
Fuel Electric
Atmosphere Vacuum/Nitrogen/Hydrogen
Effective Chamber Dimensions 500*500*650mm(W*H*D)
Transport Package Wooden Packaging
Specification 1650*1950*2000mm(W*H*D)
Trademark Chitherm
Origin China
HS Code 8514101000
Production Capacity 50 Sets/Year
Product Description

Electric Vacuum Furnace Chitherm Hwf160-10nh with 1650*1950*2000mm W*H*D and Advanced Technology

Application

This furnace is mainly designed for soldering process of metal parts and HTCC package, also can be used in other processes such as decarburization, annealing, quenching, scorching, hardening, thermal decomposition.

Technical Specification
  • Rated Temp.: 900ºC
  • Max Temp.: 1100ºC
  • Atmosphere: N2, H2, with humidification
  • Oxygen Content: Measured by O2 content analyzer
  • Muffle Material: SUS 310S
  • Muffle Dimensions: φ750mm×900mm(d×D)
  • Effective Chamber Dimensions: 500×500×650mm(W×H×D)
  • Loading Method: Manual, through horizontal guideway shelves
  • Heating Method: Resistance wire pre-built heating plate
  • Cooling Method: Air cooling with blower
  • Temp. Control Stability: ± 1ºC(controller with P.I.D function)
  • Profile Steps: Default 20 steps
  • Temp. Uniformity: ± 5ºC
  • Thermcouple: Type K
  • Temp Control: 3 Points
  • Temp. Measurement: 1 Point
  • Max. Heating Power: 40kW
  • Typical Insulation Power: ≤ 25kW
  • Heating Speed: ≤ 5ºC/min
  • Furnace Surface Temperature: ≤ 40ºC
  • Max Vacuum Value: ≤ 10Pa
  • Exhaust System: 1 exhaust port on the top with H2 igniter
  • Alarm System: audible and visual alarm in case of over temp, thermcouple broken, low air pressure, low water pressure, substandard vacuum level, etc
  • Weight: 1500kg
  • Appearance Colour: Light Grey
  • External Dimensions: 1650mm×1950mm×2000mm (W×H×D)
Electric Vacuum Muffle Furnace Ovens For Laboratory Annealing Decarburization 0
Delivery List
Item Note Qty.
Basic Components Furnace 1 SET
inspection certificate Certificate of major outsourced components 1 SET
Technical Documents Instructions, technical documents of the major outsourced parts, etc 1 SET
Key Components Heating Element 1 SET
Vacuum System 1 SET
Touch Screen 1 SET
Temperature Controller Azbil, CHINO or equivalent brand 1 SET
Thermocouple Type N 3 PCs
PLC SIEMENS or equivalent brand 1 SET
Spare Parts SSR 1 PC
Sealing ring 1 PC
Facility Requirements
  • Environmental Conditions: temp.: 0~40ºC, humidity ≤ 80%RH, no corrosive gas, no strong airflow disturbance
  • N2 Conditions: 99.999% purity, working pressure: 0.2-0.4MPa, Max gas consumption 2-6 m3/h
  • H2 Conditions: working pressure: 0.2-0.4MPa, Max gas consumption 1-3m3/h
  • Compressed Air Condition: dry, clean, oil-free, working pressure: 0.4-0.8MPa
  • Water Conditions: clean, no corrosive water, working pressure 0.1-0.3MPa, 3-9L/min
  • Ventilation System: non-contact access to the user pumping system, pumping capacity is greater than 15m3/h
  • Ground Requirements: horizontal, no obvious vibration, load-bearing> 500 kg/m2
  • Power Condition(TBD): capacity > 55kVA, 3-phases 5-lines, voltage 220/380V, frequency 50Hz. Fire line: yellow, green, red, Zero line: blue, Ground line: yellow-green
  • Installation Site: 3000mm×2500mm×3000mm(D×W×H), with an installation area greater than 7.5m2